SEFRAM 4451 50 MHz PROGRAMMABLE PULSE GENERATOR OPERATING MANUAL

Strona 2

Operating Manual SEFRAM 4451 M4451 A00 10 Range - 10ns to (Period - 10ns off time). Resolution - Up to 6 digits, limited to 100 ps. A

Strona 3

Operating Manual SEFRAM 4451 M4451 A00 11100ns; 50ns-1.0us; 500ns-10us; 5.0us-100us; 50us-1.0ms; 500us-10ms, 5 ms – 100 ms. Resolution -

Strona 4

Operating Manual SEFRAM 4451 M4451 A00 12 GPIB PROGRAMMING Internal - IEEE-488.2 and SCPI compatible. Address - 0-30 front panel se

Strona 5 - Safety Summary

Operating Manual SEFRAM 4451 M4451 A00 13

Strona 6

Operating Manual SEFRAM 4451 M4451 A00 14 Section 2 Installation 2.1 Introduction This section contains installation infor

Strona 7

Operating Manual SEFRAM 4451 M4451 A00 15 2.6 Grounding Requirements To minimize shock hazard, the instrument chassis and cabinet must be

Strona 8 - General Information

Operating Manual SEFRAM 4451 M4451 A00 16 2.7.2 Impedance Matching A mismatch, or different impedance in a transmission line, generates a

Strona 9

Operating Manual SEFRAM 4451 M4451 A00 17 For example: to match a 50Ω system to a 125Ω system, Z1 equals 50Ω and Z2 equals 125Ω Therefore:

Strona 10 - : - ± 1% ± 25 mV

Operating Manual SEFRAM 4451 M4451 A00 18than the combined rise times of the testing system, determine the actual rise time of the device und

Strona 11

Operating Manual SEFRAM 4451 M4451 A00 19DB-9 pin Name Note Computer DB25 pin Direction 1 2 3 4 5 6 7 8 9 N/C RXD TXD N/C GND N/C RTS CRS

Strona 12

Operating Manual SEFRAM 4451 M4451 A00 2

Strona 13 - M4451 A00

Operating Manual SEFRAM 4451 M4451 A00 20

Strona 14 - Installation

Section 3 Operating Instructions 3.1 General Description This section describes the displays, controls and connectors of the SEFRAM 4451 Pulse

Strona 15

Operating Manual SEFRAM 4451 M4451 A00 22output, parameter or mode displayed selected. 3.3 Front Panel Controls The front-panel controls

Strona 16 - RZZZ1221=−()

Operating Manual SEFRAM 4451 M4451 A00 233.6 MENU Keys These keys select the main menus for displaying or changing a parameter, function o

Strona 17 - R1 125 125 50 96 8=−=()

Operating Manual SEFRAM 4451 M4451 A00 24 F3: DELAY - This parameter is used in two instances. The first is to set the delay of the puls

Strona 18

Operating Manual SEFRAM 4451 M4451 A00 25 OUTPUT Menu F1: HILVL - Selects the pulse high level voltage. F2: LOLVL - Selects

Strona 19 - 2.10 RS-232 Configuration

Operating Manual SEFRAM 4451 M4451 A00 26 3.6.3 PULSE Menu PULSE Menu F1: RISE - Selects the pulse Rise

Strona 20

Operating Manual SEFRAM 4451 M4451 A00 27 3.6.4 MODE Menu Selects the output mode: CONT (Continuous), TRIG (Triggered), GATE (Gated), BRS

Strona 21 - Operating Instructions

Operating Manual SEFRAM 4451 M4451 A00 28 Trigger Menu F1: MAN - Selects manual as the trigger source. Pressing the M

Strona 22

Operating Manual SEFRAM 4451 M4451 A00 29 SETUPS Menu F1: RECALL

Strona 23

Operating Manual SEFRAM 4451 M4451 A00 3 INDEX Page SAFETY SUMMARY Section 1 General Information 5 1.1 Int

Strona 24

Operating Manual SEFRAM 4451 M4451 A00 30 GPIB Menu F2: RS232 -Selects the RS232 remote control mo

Strona 25

Operating Manual SEFRAM 4451 M4451 A00 31When the pulse generator finishes the diagnostic self-test routine, it enters the local state (LOGS)

Strona 26

Operating Manual SEFRAM 4451 M4451 A00 32 Error messages for SEFRAM 4451 Message Text Cause Setting conflict Can&apo

Strona 27

Operating Manual SEFRAM 4451 M4451 A00 33 SETTLING TIME is the interval required for the pulse level to enter and remain in the specified le

Strona 28

Operating Manual SEFRAM 4451 M4451 A00 34 (0.99 * Trig Rate) must be > (Period * Burst Count)

Strona 29

Operating Manual SEFRAM 4451 M4451 A00 35

Strona 30

Operating Manual SEFRAM 4451 M4451 A00 36

Strona 31

Section 4 Programming 4.1 OVERVIEW This section provides detailed information on programming the SEFRAM 4451 via the IEEE 488 bus (GPIB - General Pu

Strona 32

Operating Manual SEFRAM 4451 M4451 A00 36Setting the device to address 31 puts it in the 'off-bus' state. In this state it will not

Strona 33

Operating Manual SEFRAM 4451 M4451 A00 37 4.9.1 General Command Structure The device commands are generally defined by the SCPI standard, w

Strona 34

Operating Manual SEFRAM 4451 M4451 A00 4Section 5 Performance Check Procedure 66 5.1 Introduction 67 5.2 Test Equipment 67 5.3 E

Strona 35

Operating Manual SEFRAM 4451 M4451 A00 38The Program Data represent the values of the parameters being set, for example, the '1US'

Strona 36

Operating Manual SEFRAM 4451 M4451 A00 39 SOURCE:FREQUENCY 3KHZ;:OUTPUT:STATE ON Common Commands may be inserted in the Program Message wi

Strona 37 - Programming

Operating Manual SEFRAM 4451 M4451 A00 40Bit 3: Device-Specific Error (DDE). This bit is set to indicate that a device operation did not exe

Strona 38 - 4.9 COMMAND SYNTAX

Operating Manual SEFRAM 4451 M4451 A00 41 4.10.5.1 Command Errors A command error is in the range -199 to -100, and indicates that a syntax

Strona 39

Operating Manual SEFRAM 4451 M4451 A00 42A query error indicates that the output queue control has detected a problem. This could occur if ei

Strona 40

Operating Manual SEFRAM 4451 M4451 A00 43 4.11.3 Synchronization Commands 4.11.3.1 *OPC - Operation Complete Command The operation complet

Strona 41

Operating Manual SEFRAM 4451 M4451 A00 44COMMAND TYPE: Common Command or Query Syntax: *ESR? Response: <NR1> 4.11.4.4 *PSC - Power-On

Strona 42

Operating Manual SEFRAM 4451 M4451 A00 45Syntax: *TRG 4.11.6 Stored Settings Commands 4.11.6.1 *RCL - Recall Instrument State This comman

Strona 43

Operating Manual SEFRAM 4451 M4451 A00 46 :DOUBle [:STATe] <Boolean> :DELay <NRf> :TRANsition [:LEADing] <NRf>

Strona 44

Operating Manual SEFRAM 4451 M4451 A00 47This command is used to set the low level of the pulse. COMMAND TYPE: Setting or Query SETTING Synt

Strona 45

Operating Manual SEFRAM 4451 M4451 A00 5 Safety Summary The following safety precautions apply to both operating and maintenance personnel an

Strona 46

Operating Manual SEFRAM 4451 M4451 A00 48QUERY Syntax: [:SOURce]:VOLTage[:LEVel][:IMMediate]:PLOW? Examples: :VOLT:PLOW? Response: NRf 4.12.1

Strona 47

Operating Manual SEFRAM 4451 M4451 A00 49QUERY Syntax: [:SOURce]:VOLTage:LIMit:LOW? Examples: :VOLT:LIM:LOW? Response: NRf CONSIDERATIONS:

Strona 48

Operating Manual SEFRAM 4451 M4451 A00 504.12.1.11 Pulse Delay This command is used to set the delay from the trigger signal to the start of

Strona 49

Operating Manual SEFRAM 4451 M4451 A00 51SETTING Syntax: [:SOURce]:PULSe:HOLD<ws><option> Arguments: Type: Character Options: WID

Strona 50

Operating Manual SEFRAM 4451 M4451 A00 52This command is used to set the delay of the second pulse, from the time of the trigger, in the doub

Strona 51

Operating Manual SEFRAM 4451 M4451 A00 534.12.1.18 Trailing Edge Time This command is used to set the value of the trailing edge time. If th

Strona 52

Operating Manual SEFRAM 4451 M4451 A00 54This command is used to control the polarity of the pulse, which may be normal or complemented. The

Strona 53

Operating Manual SEFRAM 4451 M4451 A00 55:BURSt <NRf> :SOURce <MANual> | INTernal | EXTernal | BUS :TIMer <NRf> :LEVel <

Strona 54

Operating Manual SEFRAM 4451 M4451 A00 56QUERY Syntax: :TRIGger:BURSt? Response: NRf Examples: :TRIG:BURSt? 4.12.3.4 Internal Trigger Rate S

Strona 55

Operating Manual SEFRAM 4451 M4451 A00 57Syntax: :TRIGger:SLOPe? Examples: :TRIG:SLOP? Response: POS | NEG 4.12.4 Status Subsystem This s

Strona 56

Operating Manual SEFRAM 4451 M4451 A00 6 WARNING: To avoid electrical shock hazard, disconnect power cord before removing covers. Refer ser

Strona 57

Operating Manual SEFRAM 4451 M4451 A00 58Communication with the instrument can be resumed only by setting the address to a suitable value fro

Strona 58

Operating Manual SEFRAM 4451 M4451 A00 59COMMAND TYPE: Setting or Query SETTING Syntax: :SYSTem:POBuffer<ws><buffer> Arguments:

Strona 59

Operating Manual SEFRAM 4451 M4451 A00 60 4.14 SCPI COMMAND TREE 4.14.1 Root Node Root[:SOURce] :OUTPut :TRIGger :STATus :SYSTem 4.14.2

Strona 60

Operating Manual SEFRAM 4451 M4451 A00 61 4.14.4 TRIGger Subsystem :TRIGger:SLOPe POS | NEG :LEVel <NRF> :TIMer <NRf>

Strona 61

Operating Manual SEFRAM 4451 M4451 A00 62:SYSTem :COMMunicate :GPIB :ADDRess <NRf> :ERRor? :SECurity [:STATe]? ON | OFF:POB

Strona 62 - 4.14 SCPI COMMAND TREE

Operating Manual SEFRAM 4451 M4451 A00 63 4.15 RS-232 Programming

Strona 63

Operating Manual SEFRAM 4451 M4451 A00 64 4.15.1 General The INSTALLATION section of this manual describes the RS-232-C connection for the i

Strona 64

Operating Manual SEFRAM 4451 M4451 A00 65

Strona 65

Operating Manual SEFRAM 4451 M4451 A00 66 Section 5 Performance Check Procedure 5.1 Introduction This section contains the test proc

Strona 66 - 4.15.2 RS-232-C Operation

Operating Manual SEFRAM 4451 M4451 A00 68The following tests are intended to verify that the SEFRAM 4451 operates properly and meets its spec

Strona 67

Operating Manual SEFRAM 4451 M4451 A00 7

Strona 68 - Section 5

Operating Manual SEFRAM 4451 M4451 A00 69 input with a 50Ω termination. Select leading and trailing times. CHECK - The Rise and

Strona 69

Operating Manual SEFRAM 4451 M4451 A00 70 To contac us : SEFRAM Instruments et Systèmes 32, rue E. MARTEL BP55 F 42009 – SAINT-ETIENNE ced

Strona 70

Operating Manual SEFRAM 4451 M4451 A00 71DECLARATION OF CE CONFORMITY according to EEC directives and NF EN 45014 norm DECLARATION DE CONFORM

Strona 71 - (0,15€TTC/mn)

Section 1 General Information 1.1 Introduction This manual contains information required to operate, program, check and maintain the

Strona 72 - DECLARATION OF CE CONFORMITY

Operating Manual SEFRAM 4451 M4451 A00 9 1.4 SPECIFICATIONS The following specifications describe the instrument performance after a 20 mi